6.4 TROUBLESHOOTING PROCEDURES
Technical Service Manual
6 - 13
LifeCare PCA
LED
Defective numeric display PWA
Verify numeric display PWA
connection to MCU PWA
Replace numeric display PWA
Defective MCU PWA
INDICATOR
Defective icon display PWA
Verify icon display connection
to MCU PWA
Replace icon display PWA
Defective MCU PWA
KEYPAD
Defective keypad
Replace keypad
Defective MCU PWA
DOOR
Defective door lock assembly
Replace door lock assembly
Defective MCU PWA
SYRINGE
Defective mechanism
assembly
Adjust vial sensor switch
Replace mechanism assembly
Defective MCU PWA
EMPTY VIAL
Defective mechanism
assembly
Adjust vial sensor switch
Replace mechanism assembly
Defective MCU PWA
ALARM
Defective MCU PWA
ROTATION
Defective mechanism
assembly
Replace mechanism assembly
PATIENT PENDANT
ASSEMBLY
Defective patient pendant
assembly
Replace patient pendant assembly
Verify that replacement patient
pendant is color-coded blue
Table 6-3. Troubleshooting with the PVT
Test Failure
Probable Cause
Corrective Action
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...