SECTION 7 REPLACEABLE PARTS AND REPAIRS
LifeCare PCA
7 - 26
Technical Service Manual
7.3.8.12.3
Numeric
Display
PWA
Replacement
Recommended tools for this procedure are diagonal cutters and needle nose pliers.
Replacement parts for this procedure are:
PWA, Numeric Display
Post, Nylon
To replace the numeric display PWA, see
, and proceed
as follows:
1. Disconnect the infuser from AC power.
2. Separate the front and rear enclosures as described in
3. Remove the electronics assembly as described in
4. Using the diagonal cutters, snip off the ends of the four nylon hardware posts
attaching the numeric display PWA to the MCU PWA. Remove the defective numeric
display PWA.
5. Remove the remaining portion of the nylon hardware posts from the MCU PWA
using needle nose pliers.
6. Place the MCU PWA on a stable surface. Align the nylon hardware posts
on the replacement numeric display PWA with the two holes on the MCU PWA,
and press carefully to attach the numeric display PWA.
7. Reassemble the infuser in the exact reverse order of disassembly.
To verify successful replacement of the numeric display PWA, perform the PVT
in
7.3.8.12.4
Icon
Display
PWA
Replacement
Recommended tools for this procedure are diagonal cutters and needle nose pliers.
Replacement parts for this procedure are:
PWA, Icon Display
Post, Nylon
To replace the icon display PWA, see
and
, and proceed as follows:
1. Disconnect the infuser from AC power.
2. Separate the front and rear enclosures as described in
3. Remove the electronics assembly as described in
4. Using the diagonal cutters, snip off the ends of the four nylon hardware
posts attaching the icon display PWA to the MCU PWA. Remove the defective icon
display PWA.
CAUTION:
Exercise care when replacing or reinstalling the nylon
posts. Applying excessive pressure on the posts may bend or damage
the PWAs.
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...