SECTION 7 REPLACEABLE PARTS AND REPAIRS
LifeCare PCA
7 - 24
Technical Service Manual
7.3.8.12
ELECTRONICS
ASSEMBLY
COMPONENT
REPLACEMENT
Electronics assembly component replacement includes the replacement of the LCD shield,
LCD assembly, numeric display PWA, icon display PWA, and lithium coin cell battery.
To replace the electronics assembly components, see
and
and proceed as detailed in the following sections.
Figure 7-11. Electronics Assembly Components
7.3.8.12.1
LCD
Shield
Replacement
Recommended tools and materials for this procedure are an X-acto knife, isopropyl
alcohol, and a lint-free cloth.
The replacement part for this procedure is:
Shield, LCD
To replace the LCD shield, see
, and proceed as follows:
1. Disconnect the infuser from AC power.
2. Separate the front and rear enclosures as described in
3. Remove the electronics assembly as described in
4. Using the X-acto knife, lift the LCD shield from the LCD assembly.
MCU PWA
NUMERIC DISPLAY PWA
1/4 in. (6 mm) NYLON POST (4)
LCD ASSEMBLY
5/16 in. (8 mm) NYLON POST (4)
LCD SHIELD
ICON DISPLAY PWA
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...