4.5 FUSE INFORMATION
Technical Service Manual
4 - 3
LifeCare PCA
4.5
FUSE
INFORMATION
Instead of the fuse replacement part specified in
Section 7.3.9.1, Fuse Replacement
and
listed as item 37 in
, the following applies:
Fuse, 0.5 A, 250 V, Slo-Blo
Note:
The replacement procedures provided in
Section 7.3.9.1, Fuse Replacement
still apply. The only difference for an upgraded pump is the replacement
part.
4.6
OCCLUSION
TEST
Instead of the instructions referenced in
, perform the test
as described below.
Note:
To perform the occlusion test as a
stand-alone
test, use setup
Steps
1
through
25
Note:
Ensure that performance testing for occlusion pressure uses either only
new vials or vials that are not stored with water in them.
Note:
Ensure that the infuser and DPM are at the same height to obtain accurate
readings on the DPM.
To perform the occlusion test, proceed as follows:
1. Connect the distal end of the LifeCare PCA set to the DPM through a three-way
stopcock. Set the three-way stopcock to air.
2. Verify the DPM reads zero.
3. Verify the
PRESS START BUTTON TO INFUSE
message displays, then press
[START/
PAUSE]
.
4. Verify fluid discharge at the end of the stopcock, then close the stopcock.
5. Verify the infuser sounds an alarm and displays
OCCLUSION
above a message with
instructions for clearing the alarm.
6. Press the
[START/PAUSE]
touchswitch and verify the DPM reads 15.0 +/- 5.0 psi.
7. Open the stopcock to clear the occlusion alarm.
8. Remove the vial from the infuser, then press
[ON/OFF]
to power off the infuser.
4.7
OCCLUSION
ALARM
SWITCH
TEST
AND
ADJUSTMENT
Instead of the instructions referenced in
Section 7.2.5, Occlusion Alarm Switch
, follow the instructions provided in the sections below.
The recommended tools and materials for the following procedures are a No. 0 Phillips
screwdriver, water-filled syringe, DPM, three-way stopcock, and an IV administration set.
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...