SECTION 5 MAINTENANCE AND SERVICE TESTS
LifeCare PCA
5 - 28
Technical Service Manual
5.5.6
POLE
CLAMP
ASSEMBLY
INSPECTION
AND
TEST
Inspect the pole clamp assembly at least once every 12 months.
To inspect the pole clamp, see
, and proceed as follows:
1. Place the infuser on a flat, stable surface.
2. Confirm that the pole clamp assembly is securely attached to the rear enclosure.
Tighten the assembly screws if they are loose
.
3. Confirm that the rubber pad is present on the inside surface of the pole clamp
assembly. If the pad is missing, replace the assembly
.
4. Insert the LifeCare PCA key and unlock the door for unlocking the pole clamp
assembly.
5. Tighten and loosen the pole clamp shaft by rotating the pole clamp knob so that
the shaft moves through its entire length. Confirm that the shaft moves smoothly
and does not bind along its length. Replace the pole clamp assembly if it binds
or is difficult to tighten. If the pole clamp continues to bind after replacement,
replace the assembly.
6. Mount the infuser on an ICU Medical-approved IV pole and fully tighten the clamp.
Ensure that the infuser is held firmly and does not slide on the IV pole. If the clamp
does not hold the infuser securely, replace the pole clamp assembly.
7. Close and lock the security door. Attempt to turn the pole clamp knob
counterclockwise to ensure that the pole clamp locks securely in place when the
door is closed and locked.
Figure 5-12. Pole Clamp Assembly
POLE CLAMP
ASSEMBLY
8-32 x 3/4
FLAT HEAD SCREW
(4)
REAR
ENCLOSURE
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...