7.3 REPLACEMENT PROCEDURES
Technical Service Manual
7 - 33
LifeCare PCA
5. Remove the fuses and replace them with approved fuses only
.
Do not use any other fuse types.
6. Insert the fuseholder into the receptacle, then press the fuseholder down until
it clicks into position.
7. Replace the AC power cord.
8. Reinstall the battery door with the screw that was removed in Step 2.
9. Connect the infuser to AC power.
Fuse replacement is routine maintenance and no verification procedure is normally
required. However, if the infuser may have been damaged during this procedure,
perform the PVT in
7.3.9.2
POLE
CLAMP
ASSEMBLY
REPLACEMENT
The recommended tool for this procedure is a No. 2 Phillips screwdriver.
Replacement parts for this procedure are:
Assembly, Pole Clamp
Screw, 8-32 x 3/8, Flat Head, Phillips
Washer, Lock, #8
To replace the pole clamp assembly, see
, and proceed as follows:
1. Disconnect the infuser from AC power.
2. Place the infuser face down on a soft surface.
3. Using the Phillips screwdriver, remove the four 8-32 screws and lock washers.
4. Remove the pole clamp assembly.
5. Install the replacement pole clamp assembly using the screws and lock washers
that were removed in Step 3.
6. Connect the infuser to AC power.
Replacement of the pole clamp assembly is routine maintenance and no verification
procedure is normally required. However, if the infuser may have been damaged during
this procedure, perform the PVT in
.
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...