1.10 INFUSION SYSTEM INSTALLATION
Technical Service Manual
1 - 7
LifeCare PCA
1.10.3
SELF
TEST
If an alarm condition occurs during the self test, cycle the power and repeat the self test.
If the alarm condition recurs, note the message and take corrective action
(
see Section 6
)
.
Repeat the self test. If the alarm condition continues to recur, remove the infuser from
service and
contact ICU Medical
.
Note:
Do not place the infuser in service if the battery is not fully charged. To make
certain the battery is fully charged, connect the infuser to AC power for 16 hours.
Note:
When plugging the device into an AC power outlet, grasp the AC power cord
plug and use a forward motion into the socket. Do not use a sideways motion. When
unplugging the device, grasp the AC power cord plug and pull straight out. Do not
pull out using the power cord cable and do not pull out at an angle.
To perform the self test, see
Figure 1-1
, and proceed as follows:
1. Unlock and open the security door.
2. Plug the AC power cord into a grounded hospital grade, 120 V
AC
, 50-60 Hz
receptacle. Verify the AC plug indicator on the front panel is illuminated.
Note:
If the quality of earth grounding source is in doubt, use battery power.
Note:
When the AC plug indicator is off and the battery indicator on the display
is flashing, it is an indication that the infuser is operating on low battery
power and should be recharged.
3. Press
[ON/OFF]
to turn on the power. The infuser will perform a self test verifying
the integrity of the software, memory, and selected electronic functions.
4. Verify the time and date. To set the time and date, see
Section 1.10.4
.
Figure 1-1. LifeCare PCA Self Test Screens
Note:
If the infuser is running 7.02 software or higher, the Software Version screen
will display instead of the Self Test Complete screen, and will automatically transition
to the next screen after 5 seconds. Select SETTINGS> on this screen to view the Select
Setting to Change menu and set the time and date.
CAUTION:
Do not place the infuser in service if the self test fails.
LifeCare PCA
SELF TEST
LifeCare PCA
SELF TEST
SOFTWARE
07.00.00.019
DRUG LIBRARY
DEFAULT
TIME: 11:13 AM
DATE: 07/21/14
SYSTEM
SETTINGS
CONTINUE
LifeCare PCA
SELF TEST
RAM.............OK
FLASH..........OK
CPU ID.........OK
CPU..............OK
TIMER......... OK
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...