SECTION 5 MAINTENANCE AND SERVICE TESTS
LifeCare PCA
5 - 38
Technical Service Manual
5.6.5
BIOMED
MODE
TESTS
provides a list of system tests available in the
Biomed Mode
. Each test displays
prompting information on the LCD screen to serve as a guide through the test.
Note:
The battery and patient pendant must be connected to the infusion system
to complete the Biomed mode tests.
Note:
Verify the patient pendant has a label that indicates it is appropriate for
use with the LifeCare PCA pump.
No
Check Printer
alarm will sound if problems are encountered while printing
in the Biomed Mode.
To enter the Biomed Mode, proceed as follows:
1. Confirm the AC power indicator on the front panel is lit.
2. Unlock and open the security door.
3. Press and hold
[ENTER]
and
[ON/OFF]
simultaneously.
4. Verify the
BIOMED MODE
screen displays.
5. Verify the infusion system is not connected to a patient and press
[CONTINUE]
.
6. After the power up RAM test is completed, the infuser displays the
Biomed Mode
Main Menu
screen. Press the
[MAINTENANCE]
softkey.
7. Press the
[SYSTEM TESTS]
softkey.
8. Press the corresponding softkey to select the test to be performed (see
in
). Use the
[NEXT]
and
[PREVIOUS]
softkeys to navigate through
the Biomed mode tests and follow the on-screen instructions.
9. When all tests are completed, press
[EXIT]
to return to the
BIOMED MODE
screen,
then press
[ON/OFF]
to power off the infuser.
Table 5-7. System Tests
Test
Description
RAM
Test is performed to determine the integrity of the address lines used to select
specific memory locations for data access
Microcomputer RAM memory is tested
FLASH
Checksum is calculated and compared to a pre-stored and pre-calculated value
to assure the operating program residing in Flash memory is intact
CPU
Test routine is performed and the test results are verified to assure
the microcomputer is executing instructions properly
CLOCK
Real-time clock is tested to verify its accuracy
LCD
Non-displaying test pattern is written to the LCD RAM and compared with a stored
pattern to assure that the LCD screen can accurately display data
BACKLIGHT
Display backlight is turned off and on
LED
Pattern of 1s through 9s is displayed by the LEDs to verify operation
INDICATOR
Test verifies battery and AC symbols are turned on and off when the power cord
is connected and disconnected
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...