SECTION 6 TROUBLESHOOTING
LifeCare PCA
6 - 4
Technical Service Manual
CHECK
INJECTOR
The vial is detected and the
injector is not detected
Properly insert injector into
injector flange retainer or
remove syringe
NONE
CHECK
SETTINGS -
INCOMPLETE
PROGRAM
Door secured on any screen
other than the Press Start to
Begin or Setup/Change Menu
screen
or
No key pressed for two minutes
Open the door, check therapy
settings, and finish programing
therapy settings
1 minute
CHECK
SETTINGS -
LOADING DOSE /
PRIME NOT
COMPLETED
A Loading Dose or Prime
operation is waiting for operator
input to proceed
Dismiss the alarm by pressing
any key, then follow instructions
on the screen to complete the
operation
NONE
CHECK SYRINGE
Syringe (vial and injector)
is not properly loaded
Properly insert syringe into
cradle assembly and injector
flange retainer
Close tubing slide clamp
immediately
NONE
CHECK VIAL
Injector is detected and vial
is not properly loaded
Properly insert vial into cradle
assembly or remove injector
from injector flange retainer
Close tubing slide clamp when
handling vial
NONE
DEAD BATTERY
Battery has been completely
discharged
Connect to AC power
Plug infuser into power outlet or
battery may be damaged
NONE
DECIMAL POINT
NOT ALLOWED!
Attempting to select the decimal
point button when programming
in micrograms (mcg)
Press
[CLR]
and enter valid
whole number
N/A
DOOR OPEN
Door left open or unlocked
for more than two minutes
Close and lock the door
NONE
EMPTY SYRINGE
Empty vial is detected
Press
[AUDIO PAUSE]
to
silence the alarm
Press
[STOP]
Close slide clamp immediately
Unlock and open the security
door and squeeze the cradle
release mechanism to replace
vial
Infuser may be turned off while
preparing new vial
2 minutes
INFUSER
STOPPED
Infuser paused for more than
two minutes
Close tubing slide clamp when
handling vial
Press
[START/STOP]
to
resume infusion, or unlock door
2 minutes
Table 6-1. Operational Alarm Messages
Alarm Name
Possible Cause
Corrective Action
Audio Pause
Duration
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...