Technical Service Manual
A - 1
LifeCare PCA
APPENDIX
USE
OF
THE
INFUSION
SYSTEM
IN ELECTROMAGNETIC
ENVIRONMENTS
The infusion system is intended for use in the electromagnetic environment specified in
,
, and
. The user of the infusion system should
assure that it is used only in the appropriate environment.
ELECTROMAGNETIC
EMISSIONS
details electromagnetic emissions compliance and guidance.
Table A-1.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
Emissions Test
Compliance
Electromagnetic Enforcement - Guidance
RF Emissions
CISPR11
Class B
The infuser is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low voltage power supply network
that supplies buildings used for domestic purposes
Harmonic Emissions
IEC 61000-3-2
Class B
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies