SECTION 5 MAINTENANCE AND SERVICE TESTS
LifeCare PCA
5 - 8
Technical Service Manual
Figure 5-4. Front View - Cleaning the Vial Door
4. Use a brush to remove debris and then moist swabs to clean any openings or hard
to reach areas in the highlighted region in
.
Note:
Do not allow cleaning fluid to flow into door lock opening or Vial Cradle
track opening.
5.2.6
CLEANING
THE
VIAL
COMPARTMENT
Spilled fluid that is not cleaned properly and promptly at the patient site may accumulate
and dry. If sticky or high-viscosity fluids are spilled in the vial compartment, clean the
infuser as soon as possible. Dried, built-up residue from these types of fluids can affect
the operation of the vial sensor pin, injector sensor pin, and cradle mechanism. Dried fluid
on the injector flange retainer can also affect the way the injector fits. Fluid or contaminates
on the infuser's barcode reader window can prevent the barcode reader from operating.
The precision mechanisms located behind the vial door include parts and sensors that
can be damaged by rough handling and fluid ingress. For this reason, the components in
the vial compartment
should only be cleaned as follows:
1. When cleaning the bar code reader, use recommended CleanTex CT 811 OpticPad
Optic Surface Cleaner, or a soft, lint-free cloth moistened with a streak-free optical
glass cleaner.
CAUTION:
TO AVOID FLUID INGRESS, DO NOT SPRAY FLUIDS OR
CLEANING SOLUTIONS ON THE BAR CODE READER WINDOW AND PRESS
LIGHTLY ON THE BAR CODE READER WINDOW WHILE CLEANING.
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...