5.5 PREVENTIVE MAINTENANCE
Technical Service Manual
5 - 19
LifeCare PCA
5.5
PREVENTIVE
MAINTENANCE
ICU Medical requires that preventive maintenance (PM) be performed at least once every
12 months
. Replace components as required by visual inspection and test results.
Complete the checklist in
.
• The sealed, lead-acid battery must be replaced at least once every
24 months
.
• The mechanism assembly must be replaced at least once every
60 months
.
• The lithium coin cell battery must be replaced at least once every
120 months
.
• Perform the Performance Verification Test (PVT) at least once every
12 months
along
with the visual inspections.
Perform the preventive maintenance inspections and tests according to the following
steps:
1.
Section 5.5.1, PM/PVT Checklist
2.
Section 5.5.2, Labels Inspection
3.
Section 5.5.3, AC Power Cord Inspection and Test
4.
Section 5.5.4, Front Enclosure, Rear Enclosure, Cradle Assembly, and Security
Door Inspection and Test
5.
Section 5.5.5, Rubber Foot Pad Inspection
6.
Section 5.5.6, Pole Clamp Assembly Inspection and Test
7.
Section 5.5.7, Keypad, Displays (LED/LCD), and Indicators Inspection
8.
Section 5.5.8, Patient Pendant Inspection
9.
Section 5.5.9, Barcode Reader Window Inspection, Test, and Cleaning
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...