5.5 PREVENTIVE MAINTENANCE
Technical Service Manual
5 - 29
LifeCare PCA
5.5.7
KEYPAD,
DISPLAYS
(LED/LCD),
AND
INDICATORS
INSPECTION
Inspect the keypad, displays (LED/LCD), and indicators at least once every 12 months.
To inspect the keypad, displays (LED/LCD), and indicators, see
,
and proceed
as follows:
1. Place the infuser on a flat, stable surface.
2. Inspect the keypad for tears, cracks, or other damage. Replace the keypad if damage
is observed
.
3. Inspect the keypad for worn or illegible numbers or letters. Replace the keypad
if letters or numbers are not readable.
4. Confirm that the keypad has retained its color. Replace the keypad if any of the
keys have faded or worn color.
5. Inspect the LED display area for cracks or other damage. Confirm that the display
is clean. If the display appears to be damaged, replace the LED display
.
6. Inspect the LCD display area for cracks or other damage. Confirm that the display
is clean. If the display appears to be damaged, replace the LCD
.
To test the keypad, displays (LED/LCD), and indicators, access the
System Tests
in
Biomed Mode
.
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...