5.2 CLEANING
Technical Service Manual
5 - 5
LifeCare PCA
To clean the exterior surfaces of the infuser:
1. Rest the infuser on one side to expose the bottom
(
)
.
Figure 5-1. Bottom View - Cleaning the Bottom of the Infuser
2. Use a brush to remove debris and then moist swabs to clean any openings or hard
to reach areas highlighted in
3. Using lint-free cloths dampened in the approved cleaning solution or wipes, wipe
the bottom surface.
4. Lift the infuser.Using lint-free cloth dampened in the approved cleaning solution or
wipes, wipe the surface where it was resting and place it upright.
5. Using a lint-free cloth dampened in the approved cleaning solution or wipes, wipe
the top and side surfaces. Use a combination of brush and swab as needed to clean
the seam that joins the front and rear parts of the infuser case highlighted in
.
Note:
Do not allow fluid into any ports or vents on the infuser.