SECTION 5 MAINTENANCE AND SERVICE TESTS
LifeCare PCA
5 - 42
Technical Service Manual
2. Connect the safety analyzer ground lead to the device equipotential post.
3. Check the leakage current with the safety analyzer. Leakage current must not
exceed specifications in
4. Measure the resistance of the AC connector ground lug with the safety analyzer.
Resistance should not exceed specifications in
Note:
The door hinge is not grounded.
5. Connect the device to AC power and confirm the AC indicator light is lit.
NC = Normal Condition; SFC = Single-Fault Condition
5.6.9
END
OF
THE
PVT
If any tests fail, see
or
Note:
If the Stuck Injector Switch Pin Test in section
fails, perform
the procedure for cleaning the injector switch pin in
repeat the test. If the test fails after cleaning, replace the mechanism
assembly as described in
or
If all Performance Verification Tests have been successful, proceed as follows:
1. Clear device history.
2. Reset the LifeCare PCA infusion system to the original configuration.
3. Verify the correct time and date.
4. Return the LifeCare PCA infusion system to service.
Table 5-8. Electrical Safety Measurements
Measurement
Condition
Neutral
Switch
Polarity
Switch
Leakage
Switch
Not to
Exceed
Touch leakage current
NC - ground
intact
Closed
Normal
Default (neither
Chassis nor
Earth)
100
A
Touch leakage current
SFC - open
neutral
Open
Normal
Default (neither
Chassis nor
Earth)
500
A
Touch leakage current
SFC - open
ground
Closed
Normal
Chassis
500
A
Earth leakage current
NC
Closed
Normal
Earth
500
A
Earth leakage current
SFC - open
neutral
Open
Normal
Earth
1000
A
Chassis ground resistance
Cord connected
N/A
N/A
N/A
0.2
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...