7.3 REPLACEMENT PROCEDURES
Technical Service Manual
7 - 29
LifeCare PCA
7.3.8.14
MECHANISM
ASSEMBLY
REPLACEMENT
Recommended tools for this procedure are a No. 2 Phillips screwdriver and 5/16 inch
nutdriver.
Replacement parts for this procedure are:
Assembly, Mechanism
Screw, 10-32 x 1, Hex Head
Screw, 10-32 x 1 3/4, Hex Head
To replace the mechanism assembly, see
, and proceed as follows:
1. Disconnect the infuser from AC power.
2. Separate the front and rear enclosures as described in
3. Remove the cradle assembly as described in
4. Remove the mechanism lever arm as described in
.
5. Disconnect the three mechanism cables from the MCU PWA.
6. Place the infuser face down on a soft surface. Using the nutdriver, remove the three
10-32 screws that secure the mechanism assembly to the front enclosure.
7. Install the replacement mechanism assembly using the screws that were removed
in Step 6.
8. Reinstall the mechanism lever arm and cradle assembly.
9. Reassemble the infuser in the exact reverse order of disassembly.
Note:
Verify all connectors are fully seated before the front and rear enclosures
are closed and fastened.
To verify successful replacement of the mechanism assembly, perform the PVT
in
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...