7.3 REPLACEMENT PROCEDURES
Technical Service Manual
7 - 15
LifeCare PCA
7.3.8.3
DOOR
LATCH
HOOK
REPLACEMENT
The recommended tool for this procedure is a No. 1 Phillips screwdriver.
Replacement parts for this procedure are:
Hook, Door Latch
Screw, 4-40 x 3/16, Flat Head, Phillips
To replace the door latch hook, see
, and proceed as follows:
1. Disconnect the infuser from AC power.
2. Unlock and open the security door.
3. From inside the door, using the Phillips screwdriver, remove the 4-40 flat head
screw that attaches the latch hook to the door.
4. Install the replacement latch hook from inside the door using the screw
that was removed in Step 3.
5. Close and lock the security door.
Replacement of the door latch hook is routine maintenance and no verification procedure
is normally required. However, if the infuser may have been damaged during
this procedure, perform the PVT in
.
7.3.8.4
CRADLE
ASSEMBLY
REPLACEMENT
The recommended tool for this procedure is a No. 2 Phillips screwdriver.
Replacement parts for this procedure are:
Assembly, Cradle
Screw, 6-24 x 5/8, Pan Head, Phillips
To replace the cradle assembly, see
, and proceed as follows:
1. Disconnect the infuser from AC power.
2. Open the security door.
3. Place the infuser face up on a soft surface.
4. Using the Phillips screwdriver, remove the two 6-24 screws that secure the cradle
assembly and remove the cradle assembly.
5. Install the replacement cradle assembly using the screws that were removed
in Step 5.
6. Close and lock the security door.
To verify successful replacement of the cradle assembly, perform the PVT in
.
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...