SECTION 7 REPLACEABLE PARTS AND REPAIRS
LifeCare PCA
7 - 16
Technical Service Manual
7.3.8.5
MECHANISM
LEVER
ARM
REPLACEMENT
The recommended tool for this procedure is a medium size flat blade screwdriver.
The replacement part for this procedure is:
Arm, Lever, Mechanism
To replace the mechanism lever arm, see
, and proceed as follows:
1. Disconnect the infuser from AC power.
2. Separate the front and rear enclosures as described in
3. Remove the cradle assembly as described in
4. Using the flat blade screwdriver, lift the mechanism lever arm off the pivot post
and pull the lever arm through the front of the infuser.
5. Install the replacement mechanism lever arm in the exact reverse order of removal.
6. Reinstall the cradle assembly.
7. Reassemble the infuser in the exact reverse order of disassembly.
To verify successful replacement of the mechanism lever arm, perform the PVT
in
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...