SECTION 5 MAINTENANCE AND SERVICE TESTS
LifeCare PCA
5 - 14
Technical Service Manual
5.3.2
DISINFECTION
SUPPLIES
To disinfect the infuser, use clean, soft, lint-free cloths moistened with an approved
disinfectant
(
or disposable commercial wipes.
To disinfect hard-to-reach areas, use cotton-tipped or foam-tipped swabs. (If using
cotton-tipped swabs, make sure they do not leave a fiber residue.)
5.3.3
PREPARING
THE
INFUSER
FOR
DISINFECTION
Before you begin the disinfection process, perform the following:
1. Unplug the infuser.
2. Ensure that all visible soil was removed during disinfection and that the infuser
and accessories are completely dry.
3. Inspect enclosure for cracks and damage. If cracked or damaged, disinfect and
replace the enclosure
)
.
5.3.4
GENERAL
DISINFECTION
INSTRUCTIONS
This section describes the methods you should use when disinfecting the infuser. For
specific instructions on disinfecting each part of the infuser using these methods, see
.
Table 5-5. Disinfection Methods
Type of Disinfection Instrument
Method
Lint-Free Cloth
1.
Wearing gloves, dampen the cloth with the
approved disinfecting solution (
) and wipe in an outward spiral pattern
on the desired surface(s).
2.
When the cloth becomes soiled, start wiping
with the unsoiled part of the cloth. Replace
the cloth as needed.
Disposable Disinfectant Wipe
1.
Wearing gloves, unfold the wipe completely,
and use in an outward spiral pattern on the
desired surface(s).
2.
When the disinfectant wipe becomes soiled,
start wiping with the unsoiled part of the
disinfectant wipe. Replace the wipe as
needed.
Foam or Cotton-Tipped Swab
1.
Wearing gloves, dip a swab in an approved
disinfecting solution (
).
2.
Press the tip against the solution container
to remove excess fluid.
3.
Use the swab to clean hard to reach areas.
Change swabs as needed.
4.
Dampen a lint-free cloth with the approved
disinfecting solution to remove any fiber
residue left behind, if using a cotton swab.
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...