SECTION 7 REPLACEABLE PARTS AND REPAIRS
LifeCare PCA
7 - 38
Technical Service Manual
Figure 7-13. Rear Enclosure Harness Assembly
7.3.9.8
BATTERY
PAD
REPLACEMENT
Recommended tools and materials for this procedure are an X-acto knife, isopropyl
alcohol, and a lint-free cloth.
The replacement part for this procedure is:
Pad, Battery
To replace the battery pad, see
, and proceed as follows:
1. Disconnect the infuser from AC power.
2. Separate the front and rear enclosure assemblies as described in
3. Using the X-acto knife, carefully remove the battery pad from the rear enclosure.
4. Using a lint-free cloth moistened with isopropyl alcohol, clean any adhesive residue
from the enclosure.
5. Remove the protective backing from the replacement battery pad to expose
the adhesive and install the pad into the rear enclosure.
6. Reassemble the infuser in the exact reverse order of disassembly.
7. Connect the infuser to AC power.
To verify successful replacement of the battery pad, perform the PVT in
REAR ENCLOSURE
HARNESS ASSEMBLY
SPIRAL WRAP
TIE WRAPS
REAR ENCLOSURE
ASSEMBLY
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...