7.2 ADJUSTMENT PROCEDURES
Technical Service Manual
7 - 3
LifeCare PCA
Figure 7-1. Separating the Front and Rear Enclosure Assemblies
7.2.3
INJECTOR
SENSOR
SWITCH
ADJUSTMENT
The recommended tools and materials for this procedure are a No. 0 Phillips screwdriver
and red GLPT insulating varnish.
To adjust the injector sensor switch, see
, and proceed as follows:
1. Disconnect the infuser from AC power.
2. Separate the front and rear enclosures as described in
3. Using the Phillips screwdriver, loosen the two switch mounting screws until
the switch holder can be easily moved. Do not loosen the screws so much that
the switch bracket rocks from side to side.
4. Position the switch bracket parallel to the surface and 1/32 inch from the front
enclosure, then tighten the screws.
5. Apply red GLPT insulating varnish to the screw heads.
6. Reassemble the infuser in the exact reverse order of disassembly.
7. Close and lock the security door.
#6-32 x 1/2
PAN HEAD
SCREW (4)
#6 LOCK WASHER (3)
P9
P17
P14
REAR
ENCLOSURE
FRONT
ENCLOSURE
SPLIT LOCK
WASHER