235
_____________________________________________________________________________________________
11.1 General Specifications
______________________________________________________________________________________________
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
11.1.5 External Control Terminals
Configuration of terminals
M3-screw terminal block (12 input terminals, 16 output terminals)
The external control terminals are isolated from the 8730-10, 8731-10.
Power supply terminals
External power supply V
EXT
: between EXT_DCV and EXT_COM
Maximum external input voltage V
EXTmax
: 30 V DC
Minimum external input voltage V
EXTmin
: 5 V DC
Internal power supply V
INT
: between ISO_DCV and ISO_GND
Output voltage 5 V
0.5 V 40 mA (Total)
Input terminals
External trigger: EXT_TRIG, Measurement start: START,
Measurement stop: STOP, External Sampling: EXT_SAMP
Switching test mode: MODE_SET, MODE3, MODE2, MODE1, MODE0,
Hold: HOLD, Comparison stop: JUDGE_OFF,
Multi-function terminal: EXT_IN
Signal level: HIGH V
EXT
-1.0 to V
EXT
[V] (When there is no load)
LOW 0 to 1.0 [v]
Output terminals
Trigger output: TRIG_OUT, Area comparison 1: OK_AREA1, NG_AREA1,
Area comparison 2: OK_AREA2, NG_AREA2
Value comparison: OK_VALUE, NG_VALUE,
Comparison result being output: JUDGE_OUT, Error detected: ERROR,
Processing: BUSY, Measuring: STORAGE,
Multi-function terminal: EXT_OUT1, EXT_OUT2, EXT_OUT3, EXT_OUT4,
EXT_OUT5
Signal level: HIGH V
EXT
-1.0 to V
EXT
[V] (When there is no load)
LOW 0 to 1.0 [V]
Open collector output (pull up resistance 10 k
Ω
)
Maximum sink current 140 mA max.
Maximum voltage to earth
300 V rms or 60 V DC (between the input/output terminals and the main
unit)
Maximum input voltage
30 V DC
11.1.6 External Interface
RS-232C
In accordance with EIA RS-232C, D-sub 9 pin connector
LAN
IEEE802.3 Ethernet 10BASE-T