6.4 File Screen
______________________________________________________________________________________________
6.4.1 Loading the Data
Command:
LOAD
SAVE
CH: 1, 2
LOAD
CH: 1, 2
SAVE
CH: 1, 2
LOAD
CH: -, 1
Data saved on
PC card
Destination
channel of the
8730-10, 8731-10
In the example as
shown on the left,
CH1 data and CH2
data are saved on the
PC card.
Loads the data of CH2
on the PC card to CH1.
Clears the waveform of
CH2.
NOTE
Transfers the 8730-10, 8731-10's setting conditions, waveform data, or
area data from a PC card to the internal memory of the 8730-10, 8731-10.
When loading measurement data, the channel can be specified.
You can switch channels by using this command, for example, to read
CH1 data from the PC card to CH2 of the 8730-10, 8731-10.
Procedure
1. Press the
FILE
key to display the File
screen.
2. Use the
CURSOR
key to select a file to
load, then press the
ENTER
key.
3. When loading waveform data, specify
the destination channel.
(Note 1)
When loading area data, specify a
destination area from A01 to A16.
4. Press the
START
key to start loading.
Press the
STOP
key to cancel.
Start the loading.
Cancel the loading.
Data saved in BMP format or text format
(CSV) cannot be read by the 8730-10, 8731-
10.
Only files with the following extensions can
be read by the 8730-10, 8731-10.
MEM : Waveform data in 8730-10 (binary)
format
ARE : Area data
SET : Setting data
ROM : Index file for loading area data and
comparison condition data all at
once
(Note 1) Channel selection is not possible
with the 8730-10.
The 8835-01 waveform data measured with the
8936 ANALOG UNIT can be read (up to 500
DIV).
If the waveform data are equal to or greater
than 501 DIV, the waveform data file cannot
be loaded.