4.3 Details of Input Setting
______________________________________________________________________________________________
Increases in number
Moves the cursor to
the right
Decreases in number
Moves the cursor to the left
Confirms the value
0 0 0 . 0 0 0 0
A
B
設定方法
Value input screen
設定方法
(A)
(B)
Setting Procedure
Output of sensor
Scaling
After conversion
Upper limit: 5.78 [V]
Lower limit: 1.23 [V]
Upper limit: 10 [EU]
Lower limit:
0 [EU]
(Example)
Entering values in Variable function
Use the
CURSOR
keys and
ENTER
key to enter values. The function of
each key is shown in the figure below.
The input field is nine digits long (excluding the decimal point) as shown
in the figure below. Digits (A) and (B) are used for signs and prefixes.
The five most significant digits of numbers are effective.
(A): Displays a blank or -. A blank indicates a positive value, the -
indicates a negative value.
(B): Displays a blank or SI prefix. For details of the SI prefix, see 13.2.5.
<<Related Issues>>
Variable function and scaling function processing can be carried out
simultaneously. (See 4.3.8) For example, this is convenient when you want
to display sensor output in full span.
With the scaling function, voltage from the sensor may be converted into
the type of physical quantity you wish to measure. However, the
waveform display will not change. "To display sensor output in full span,"
set as shown below.
Lower limit
0 [EU], Upper limit
10 [EU]