![Hioki 8730-10 Скачать руководство пользователя страница 217](http://html1.mh-extra.com/html/hioki/8730-10/8730-10_instruction-manual_2137220217.webp)
201
_____________________________________________________________________________________________
8.4 Input Terminals
______________________________________________________________________________________________
8.4.5 Multi-function terminal (Input) <EXT_IN Terminal (A14)>
HOLD
T
T
T
EXT_IN
OK
、
NG
、
JUDGE_OUT
EXT_IN
Comparison clear
Clears result
Wait for
trigger
Trigger activated
When function of "
EXT_IN
"is set to "
Clear Judge-output
"
Trigger activated
Trigger activated
Operation of
the 8730-10
8731-10
Storing
waveform
Evalu
-ation
Display Wait for
trigger
Storing
waveform
Evalu
-ation
Display
Storing
waveform
Evalu
-ation
Clears result
Does not clear
comparison result
Within 50 ms
More than 60 ms
(Clears comparison output)
Starts measurement
Stops measurement
More than 60 ms
More than 60 ms
When function of "
EXT_IN
" is set to "
START&STOP
"
START & STOP
Wait for
trigger
The EXT_IN terminal (terminal No.: A14) is a multi-function terminal for
which you can select a function.
The following two functions are available:
1.
Clears comparison output:
Clears comparison result output (OK, NG, or JUDGE_OUT) at the falling
signal (active LOW) of the EXT_IN terminal. Note that output is not
cleared when the HOLD terminal (terminal No.: A15) is "L."
2.
START & STOP:
Starts measurement at the falling signal (active LOW) of the EXT_IN
terminal and stops measurement at the rising of an input signal (active
HIGH). Measurement cannot start when the BUSY terminal is LOW.
For information on setting the multi-function terminal, see 7.5.2.
The maximum delay from the falling or rising edge until activation is 50 ms.
Activated when a edge is detected.
Input Signal
The EXT_IN terminal (A14) inputs a signal to LOW or HIGH level
according to the function selected.
<When using an internal power supply>
HIGH level voltage: Max.: 5.0 [V], Min.: 4.0 [V]
LOW level voltage : Max.: 1.0 [V], Min: 0 [V]
<When using an external power supply>
External power supply voltage V
EXT
: Max.: 30 [V] DC, Min.: 5 [V] DC
HIGH level voltage : Max.: V
EXT
[V], Min.: V
EXT
-1.0 [V]
LOW level voltage : Max.:
1.0 [V],
Min.: 0 [V]
<Timing>