![Hioki 8730-10 Скачать руководство пользователя страница 213](http://html1.mh-extra.com/html/hioki/8730-10/8730-10_instruction-manual_2137220213.webp)
197
_____________________________________________________________________________________________
8.4 Input Terminals
______________________________________________________________________________________________
8.4.2 External Trigger <EXT_TRIG Terminal (A7)>
E X T _ T R IG
500μs以上
500μs以上
1ms以上
More than 500
μ
s
More than 500
μ
s
More than 1 ms
A trigger signal may be input from an external device.
The trigger is activated when the EXT_TRIG terminal (terminal No.: A7)
detects a falling signal (active LOW).
The maximum delay from the falling edge until activation is 100
μ
s.
When STORAGE terminal begins to output LOW signal, triggering is
acceptable. (See 8.5.7)
Input Signal
The EXT_TRIG terminal (A7) must be shorted with EXT_COM (A3), or a
low-level signal must be input to these terminals.
The trigger is activated at the falling edge.
Input signals of up to 1 KHz are available to activate the trigger.
<When using an internal power supply>
HIGH level voltage : Max.: 5.0 [V], Min.: 4.0 [V]
LOW level voltage : Max.: 1.0 [V], Min.: 0 [V]
<When using an external power supply>
External power supply voltage V
EXT
: Max.: 30 [V] DC, Min.: 5 [V] DC
HIGH level voltage : Max.: V
EXT
[V], Min.: V
EXT
-1.0 [V]
LOW level voltage : Max.:
1.0 [V],
Min.: 0 [V]
<Timing>