BASIC OPERATION
110
Setting: Move the cursor to the item to be changed
(it turns into a red letter on a gray background),
press Enter, and when the up and down arrow
signs appear, operate the up and down direction
keys to select the required option (Enable or
Disable), and then press Enter to confirm.
Fail mode
Used to select the fail mode of Limit 1 test.
When selecting In, if the
measurement reading
is
within the compliance limit range, press the F3 key
to display the green PASS logo; if the
measurement
reading
is restricted by the compliance value, press
the F3 key to display the red FAIL logo;
When selecting Out, if the
measurement reading
is
within the compliance limit range, press the F3 key
to display the red FAIL logo; if
measurement
reading
is restricted by the compliance value, press
the F3 key to display the green PASS logo;
Setting: Move the cursor to the setting item (it turns
into a red letter on a gray background), and press
Enter. After the up and down arrow signs appear,
operate the up and down direction buttons to select
the required option (In or Out), and then press
Enter to confirm.
CMPL
pattern
Used to set the Fail pattern of Limit 1 test:
When the Digout size is 3 bits, the value is 0-7;
when the Digout size is 4 bits, the value is 0-15.
Setting: Move the cursor to the item to be changed
(it becomes a red letter on a gray background),
press Enter, and it is in the programming state.