BASIC OPERATION
90
into a red letter with a gray background), and press
Enter. When the up and down arrow signs appear,
operate the up and down direction buttons to select
the required option (Enable or Disable), and then
press Enter to confirm.
Readback
When Readback option selects Enable, the
measurement reading are the actual output voltage
and current which will be used for resistance
calculation. When Readback option selects Disable,
the GSM uses the programmed value for resistor
calculation, which may produce incorrect
measurement values. When the source is in
compliance(limited by compliance value), selecting
Readback option as Disable can make ohms
measurements correct.
NOTE: The Readback option is Enable by
default to achieve optimal accuracy.
Setting: Move the cursor to the setting item (it turns
into a red letter with a gray background), and press
Enter. When the up and down arrow signs appear,
operate the up and down direction buttons to select
the required option (Enable or Disable), and then
press Enter to confirm.