BASIC OPERATION
102
When the Size is 4-bit, the range of the pattern value
is 0-15.
When the Size is 16-bit, the range of the pattern value
is 0-65535.
Setting: Move the cursor to the item to be
programmed (it turns into a red letter on a gray
background), and press Enter. After the up and down
arrow signs appear, operate the up and down
direction keys to select the required option (3bit, 4bit
or 16bit), and then press Enter to confirm.
Mode
Choose Grading or Sorting mode
Grading mode
: if a measurement reading has passed
the compliance test and is within all HI/LO tolerance
limits, PASS will be displayed.
Sorting fail option cannot be set in Grading mode.
When Grading is set to Immediate, the measurement
process will terminate at the first fail situation. If the
measurement reading is less than the Low limit of any
of Limit 2, 3, 5-12, the I/O port will output the
corresponding Lo_fail value. If the measurement
reading is higher than the High limit of any of Limit
2, 3, 5-12, the I/O port outputs the corresponding
Hi_fail value. If HW_Limits and SW_Limits are all
PASS, the value of Pass pattren will be outputed, and
the measurement process will be terminated.
For example, when the GSM is connected to a 20Ω
resistor for Limit test, set the GSM as I-Source,
programming Isrc to 105mA, Cmpl to 21V, and
Measure volt-range to 20V. On the main interface,
long pressing F3 button to enter the Limit setting
interface, and setting the options in Digout, HW-
Limits, SW-Limits and Pass pattern menus. After the
setting is completed, press F6 (Cancel) button to
return to the Limit test interface, press the Output