GSM-20H10 User Manual
119
State Description
REAR
Display when setting as rear-panel output. Otherwise, it is
the front-panel output
REM
Remote control
CV
When the GSM is set as I-Source, in case that the terminal
output voltage is restricted by Cmpl or Measure vol-
range, the GSM operates in CV (constant voltage) mode
When the GSM is set as V-Source, in case that the terminal
output current is not restricted by Cmpl or Measure cur-
range, the GSM operates in CV (constant voltage) mode
CC
When the GSM is set as V-Source, in case that the terminal
output current is restricted by Cmpl or Measure curr-
range, the GSM operates in CC (constant current) mode
When the GSM is set as I-Source, in case that the terminal
output voltage is not restricted by Cmpl or Measure vol-
range, the GSM operates in CC (constant current) mode
ARM
Source-measure operation is in progress
TRIG
Select external trigger source (Tlink, Rising Edge, Falling
Edge, Edge)
OVP
When the overvoltage protection function is disable, the
OVP mark is gray, when the OVP function is enable, the
OVP mark is black, and the mark is red when the OVP
function is triggered.
ERR
When reading failure or invalid calibration steps occur,
ERR mark will display in status bar. For detailed error
codes, see page 358.
PASS
Limit test PASS
FAIL
Limit test FAIL