GSM-20H10 User Manual
259
will accommodate all source values (Best Fixed
Range). You can use source auto range if sweeping
through one or more source ranges.
When the sweep is started, the source will output the
specified start level and, after the delay period of the
SDM cycle, a measurement is performed.
<n> =-210 to 210 Specify start voltage level
DEFault 0V
MINimum -210V
MAXimum +210V
The sweep continues until the source outputs the
specified stop level. At this level, the instrument
again performs another measurement (after the SDM
delay) and then stops the sweep.
The source-measure points in a sweep can be set by
specifying a step size, or by specifying the number of
source-measure points in the sweep.
:STARt and :STOP are coupled to :CENTer
and :SPAN. Thus, when start and stop values are
changed, the values for center and span are affected
as follows:
Center=(Start + Stop)/2
Span=Stop-Start
Example
:SOURce:VOLTage:STARt DEFault
Command
:SOURce[1]:VOLTage:STARt?
Function
:STARt? Query start level for sweep.
:STARt? DEFault Query *RST default level.
:STARt? MINimum Query lowest allowable level.
:STARt? MAXimum Query highest allowable
level .
Example
:SOURce:VOLTage:STARt?