GSM-20H10 User Manual
215
Function
This Command is used to define the digital output fail
patterns for the specified tests (0 to 7, 3-bit; 0 to 15, 4-
bit). NOTE that the fail patterns for Limits 2, 3, 5-12
apply only to the grading mode.
Please refer to the Command:
CALCulate2:LIMit[1]:COMPliance:SOURce2 <NRf>
|<NDN>
Example
:CALCulate2:LIMit2:UPPer:SOURce2 0
Command :CALCulate2:LIMitx:UPPer:SOURce2?
Function
Query source value for specified limit.
Example
:CALCulate2:LIMit2:UPPer:SOURce2?
Command :CALCulate2:LIMitx:PASS:SOURce2
<NRf> | <NDN>
Function
This Command is used to define the 3-bit or 4-bit
output pattern for the Digital I/O Port when a test
(limit 2, 3, 5-12) for the sorting mode passes. NOTE
that the output value can be specified in binary, octal,
decimal, or hexadecimal format. Use the table
provided in the “Description” for the :SOURce
Command to determine the parameter value for the
desired decimal digital output pattern.
<NRf> = 0 to 7 (3-bit) Decimal value
0 to 15 (4-bit) Decimal value
<NDN> = 0 to #b111 (3-bit) Binary value
0 to #b1111 (4-bit) Binary value
0 to #q7 (3-bit) Octal value
0 to #q17 (4-bit) Octal value
0 to #h7 (3-bit) Hexadecimal value
0 to #hF (4-bit) Hexadecimal value