COMMAND SYNTAX
252
With this Command, the amplitude is updated when
the GSM is triggered to perform a source-measure
operation. For example, if the instrument is waiting in
the trigger layer for an external trigger, the amplitude
of the source will not update until that external trigger
is received by the GSM.
The MINimum and MAXimum parameters are only
valid if the highest source range is presently selected.
Sending the MINimum or MAXimum parameters on a
lower source range will generate error -221 (Setting
Conflict).
<n> = -1.05 to 1.05 Set I-Source amplitude (amps)
DEFault 0A
MINimum -1.05A
MAXimum +1.05A
Example
:SOURce:CURRent:TRIGgered 0
Command :SOURce[1]:CURRent[:LEVel]:TRIGgered?
Function
:TRIGgered? Query triggered
amplitude for fixedI-Source.
:TRIGgered? DEFault Query *RST default
amplitude.
:TRIGgered? MINimum Query lowest allowable
amplitude.
:TRIGgered? MAXimum Query highest allowable
amplitude.
Example
:SOURce:CURRent:TRIGgered?
Command :SOURce[1]:VOLTage[:LEVel]:TRIGgered[:AMPLitude
] <n>
Содержание GSM-20H10
Страница 107: ...GSM 20H10 User Manual 105 The operating flow of Grading mode is shown in the figure below ...
Страница 108: ...BASIC OPERATION 106 ...
Страница 110: ...BASIC OPERATION 108 ...
Страница 166: ...SYSTEM SETTINGS 164 ...
Страница 244: ...COMMAND SYNTAX 242 Example ROUTe TERMinals ...