COMMAND SYNTAX
260
Command
:SOURce[1]:CURRent:STOP <n>
Function
This Command is used to specify the stop level for a
sweep. If using a fixed (manual) source range, the
sweep will be performed using a source range that
will accommodate all source values (Best Fixed
Range). You can use source auto range if sweeping
through one or more source ranges.
When the sweep is started, the source will output the
specified start level and, after the delay period of the
SDM cycle, a measurement is performed.
<n> = -1.05 to 1.05 Specify stop current level
DEFault 0A
MINimum -1.05A
MAXimum +1.05A
The sweep continues until the source outputs the
specified stop level. At this level, the instrument
again performs another measurement (after the SDM
delay) and then stops the sweep.
The source-measure points in a sweep can be set by
specifying a step size, or by specifying the number of
source-measure points in the sweep.
:STARt and :STOP are coupled to :CENTer
and :SPAN. Thus, when start and stop values are
changed, the values for center and span are affected
as follows:
Center=(Start+Stop)/2
Span=Stop-Start
Example
:SOURce:CURRent:STOP 0.08
Command
:SOURce[1]:CURRent:STOP?
Содержание GSM-20H10
Страница 107: ...GSM 20H10 User Manual 105 The operating flow of Grading mode is shown in the figure below ...
Страница 108: ...BASIC OPERATION 106 ...
Страница 110: ...BASIC OPERATION 108 ...
Страница 166: ...SYSTEM SETTINGS 164 ...
Страница 244: ...COMMAND SYNTAX 242 Example ROUTe TERMinals ...