GSM-20H10 User Manual
307
the measurement range will be on the same range as
the compliance setting.
<b> = OFF Disable range sync
ON Enable range sync
Example
:SENSe1:CURRent:DC:PROTection:RSYNchronize ON
Command
[:SENSe[1]]:VOLTage[:DC]:PROTection:RSYNchroni
ze <b>
Function
Turning this feature on will make the measurement
range follow the compliance range setting when
autoranging is off. When the compliance value is set,
the measurement range will be on the same range as
the compliance setting. The
[:SENSe[1]]:CURRent[:DC]:PROTection:RSYNchroni
ze <b> Command and the
[:SENSe[1]]:VOLTage[:DC]:PROTection:RSYNchroni
ze <b> Command change one setting; changing the
state of one will automatically change the other.
<b> = OFF Disable range sync
ON Enable range sync
Example
:SENSe1:VOLTage:DC:PROTection:RSYNchronize ON
Command
[:SENSe[1]]:CURRent[:DC]:PROTection:TRIPped?
Function
This Command is used to determine if the source is
in compliance. If a “1” is returned, then the source is
in compliance. A “0” indicates that the source is not
in compliance. The :CURRent:PROTection:TRIPped?
Command is used to check the compliance state of
the V-Source, and
the :VOLTage:PROTection:TRIPped? Command is
used to check the compliance state of the I-Source.
Example
:SENSe1:CURRent:DC:PROTection:TRIPped?