GSM-20H10 User Manual
271
Command
:SOURce[1]:SWEep:POINts?
Function
:POINts? Query number of sweep
points
:POINts? DEFault Query *RST default
number of sweep points
:POINts? MINimum Query lowest allowable
number of sweep points
:POINts? MAXimum Query highest allowable
number of sweep points
Example
:SOURce:SWEep:POINts?
Command
:SOURce[1]:SWEep:DIRection <name>
Function
Normally, a sweep is run from the start level to the
stop level. The :STARt and :STOP, or :CENTer
and :SPAN Commands are used to set these levels.
This Command lets you change the execution
direction of the sweep.
With DOWn selected, the sweep will begin at the
stop level and end at the start level. Selecting UP
restores sweep operation to the normal start to stop
direction.
<name> = UP Run sweep from start to stop
DOWn Run sweep from stop to start
Example
:SOURce:SWEep:DIRection UP
Command
:SOURce[1]:SWEep:DIRection?
Function
Query direction of sweep.
Example
:SOURce:SWEep:DIRection?