GSM-20H10 User Manual
253
Function
This Command performs the same as the
[:IMMediate][:AMPLitude] Command except that the
amplitude is not updated immediately.
With this Command, the amplitude is updated when
the GSM is triggered to perform a source-measure
operation. For example, if the instrument is waiting in
the trigger layer for an external trigger, the amplitude
of the source will not update until that external trigger
is received by the GSM.
The MINimum and MAXimum parameters are only
valid if the highest source range is presently selected.
Sending the MINimum or MAXimum parameters on a
lower source range will generate error -221 (Setting
Conflict).
<n> = -210 to 210 Set V-Source amplitude (volts)
DEFault 0V
MINimum -210V
MAXimum +210V
Example
:SOURce:VOLTage:TRIGgered 0
Command :SOURce[1]:VOLTage[:LEVel]:TRIGgered?
Function
:TRIGgered? Query triggered amplitude
for fixed V-Source.
:TRIGgered? DEFault Query *RST default
amplitude.
:TRIGgered? MINimum Query lowest allowable
amplitude.
:TRIGgered? MAXimum Query highest allowable
amplitude.
Example
:SOURce:VOLTage:TRIGgered?
Содержание GSM-20H10