GSM-20H10 User Manual
317
System Commands
Command
:SYSTem:PRESet
Function
This Command returns the instrument to states
optimized for front panel operation. :SYSTem:PRESet
defaults are listed in the SCPI tables
.
Example
:SYSTem:PRESet
Command
:SYSTem:POSetup <name>
Function
This Command is used to select the power-on
defaults. With RST selected, the instrument powers
up to the *RST default conditions. With PRES
selected, the instrument powers up to
the :SYStem:PRESet default conditions. Default
conditions are listed in the SCPI tables.
With the SAV0-3 parameters specified, the instrument
powers-on to the setup that is saved in the specified
location using the *SAV Command.
<name> = RST Power-up to *RST defaults
PRESet Power-up to :SYSTem:PRESet
defaults
SAV0 Power-up to setup stored at
memory location 0
SAV1 Power-up to setup stored at
memory location 1
SAV2 Power-up to setup stored at
memory location 2
SAV3 Power-up to setup stored at
memory location 3
Example
:SYSTem:POSetup SAV0
Содержание GSM-20H10