![GW Instek GSM-20H10 Скачать руководство пользователя страница 293](http://html1.mh-extra.com/html/gw-instek/gsm-20h10/gsm-20h10_user-manual_2280064293.webp)
GSM-20H10 User Manual
291
Function
When concurrent measurements are disabled, this
command is used to enable function to be measured.
The [:ON] command is used to include (enable) one
or more measurement functions in the list. NOTE
that each function specified in the list must be
enclosed in single or double quotes, and functions
must be separated by commas (,).
Examples: :FUNCtion “VOLTage”, “CURRent”
enable volts and amps functions. NOTE that there is
a stand-alone Command that can be used to enable or
disable all three measurement functions. If
concurrent measurements
are disabled, the :ON
Command can only turn on one function at a time.
<function list> = “CURRent[:DC]” Amps
measurement function
“VOLTage[:DC]” Volts
measurement function
“RESistance” Ohms
measurement function
The function selected with the SENSE:FUNC
Command will not be displayed on the front panel if
concurrent measurements are enabled.
Example
:SENSe1:FUNCtion:ON "VOLT"
Command
[:SENSe[1]]:FUNCtion[:ON]?
Function
Query functions that are enabled
Example
:SENSe1:FUNCtion:ON?
Command
[:SENSe[1]]:FUNCtion[:ON]:ALL
Содержание GSM-20H10
Страница 107: ...GSM 20H10 User Manual 105 The operating flow of Grading mode is shown in the figure below ...
Страница 108: ...BASIC OPERATION 106 ...
Страница 110: ...BASIC OPERATION 108 ...
Страница 166: ...SYSTEM SETTINGS 164 ...
Страница 244: ...COMMAND SYNTAX 242 Example ROUTe TERMinals ...