COMMAND SYNTAX
226
Display Commands
Command :DISPlay:DIGits <n>
Function
This Command is used to set the display resolution.
NOTE that you can instead use rational numbers. For
example, to select 4.5 digit resolution, you can send a
parameter value of 4.5 (instead of 5). The GSM rounds
the rational number to an integer.
<n> = 4 3.5 digit resolution
5 4.5 digit resolution
6 5.5 digit resolution
7 6.5 digit resolution
DEFault 5.5 digit resolution
MINimum 3.5 digit resolution
MAXimum 6.5 digit resolution
Example
:DISPlay:DIGits 4
Command :DISPlay:DIGits?
Function
Used to query the display resolution.
:DIGits? Query display resolution
:DIGits? DEFault Query *RST default resolution
:DIGits? MINimum Query lowest allowable display
resolution
:DIGits? MAXimum Query largest allowable display
resolution
Example
:DISPlay:DIGits?
Command :DISPlay:ENABle <b>