GSM-20H10 User Manual
331
After the buffer stores the specified number of
reading arrays (as set by the :POINTs Command), the
asterisk annuciator turns off to indicate that storage is
done.
With NEVer selected, storage into the buffer is
disabled.
<name> = NEXT Fills buffer and stops
NEVer Disables buffer storage
Example
:TRACe:FEED:CONTrol NEXT
Command
:TRACe:FEED:CONTrol?
Function
Query buffer control.
Example
:TRACe:FEED:CONTrol?
Command
:TRACe:TSTamp:FORMat <name>
Function
This Command is used to select the timestamp
format for buffer readings. With ABSolute selected,
each timestamp is referenced to the first reading
stored in the buffer. With DELTa selected,
timestamps provide the time between each buffer
reading.
<name> = ABSolute Reference to first buffer reading
DELTa Time between buffer readings
Example
:TRACe:TSTamp:FORMat ABSolute
Command
:TRACe:TSTamp:FORMat?
Function
Query timestamp format.
Example
:TRACe:TSTamp:FORMat?
Command
:TRIGger:CLEar
Содержание GSM-20H10
Страница 107: ...GSM 20H10 User Manual 105 The operating flow of Grading mode is shown in the figure below ...
Страница 108: ...BASIC OPERATION 106 ...
Страница 110: ...BASIC OPERATION 108 ...
Страница 166: ...SYSTEM SETTINGS 164 ...
Страница 244: ...COMMAND SYNTAX 242 Example ROUTe TERMinals ...