[:SENSe[1]]:CURRent[:DC]:RANGe:HOLDoff:DELay
?
Function
Query holdoff delay
Example
:SENSe1:CURRent:DC:RANGe:HOLDoff:DELay?
Command
[:SENSe[1]]:CURRent[:DC]:PROTection[:LEVel] <n>
Function
This Command is used to set compliance limits. A
current compliance limit is set for the V-Source, and a
voltage compliance limit is set for the I-Source. The
GSM cannot source levels that exceed these specified
limits. The :SENSe:CURRent:PROTection[:LIMit]
Command is used to set the current compliance for
the V-Source and the :SENSe:VOLTage:
PROTection[:LIMit] Command is used to set the
voltage compliance for the I-Source.
<n> = -1.05 to 1.05 Current compliance limit
DEFault 105uA
MINimum -1.05A
MAXimum 1.05A
Example
:SENSe1:CURRent:DC:PROTection:LEVel 1
Command
[:SENSe[1]]:CURRent[:DC]:PROTection:LEVel?
Function
:LEVel? Query compliance value.
:LEVel? DEFault Query *RST default compliance.
:LEVel? MINimum Query minimum allowable
compliance.
:LEVel? MAXimum Query maximum allowable
compliance.
Example
:SENSe1:CURRent:DC:PROTection:LEVel?