COMMAND SYNTAX
312
<b> = 0 or OFF Disable digital filter
1 or ON Enable digital filter
Example
:SENSe1:AVERage:STATe 0
Command
[:SENSe[1]]:AVERage:STATe?
Function
Query state of digital filter
Example
:SENSe1:AVERage:STATe?
Status Commands
Command
:STATus:PRESet
Function
Clears the operation event enable registers, the
measurement event enable registers and the
questionable event register, The unit will then return
to the default settings status.
When this Command is sent, the following SCPI
event registers are cleared to zero (0):
1. Operation Event Enable Register.
2. Event Enable Register.
3. Measurement Event Enable Register
Example
:STATus:PRESet
Command
:STATus:MEASurement[:EVENt]?
Function
Reads the measurement event status register.
Example
:STATus:MEASurement?
Command
:STATus:QUEStionable[:EVENt]?
Function
Read the questionable event status register.
Example
:STATus:QUEStionable?