COMMAND SYNTAX
212
Function
:UPPer? Query specified upper limit
:UPPer? DEFault Query *RST default upper limit
:UPPer? MINimum Query lowest allowable upper
limit
:UPPer? MAXimum Query largest allowable upper
limit.
Example
:CALCulate2:LIMit2:UPPer?
Command :CALCulate2:LIMit[1]:COMPliance:SOURce2 <NRf>
|<NDN>
Function
This Command is used to define the LIMIT 1 failure
pattern (0 to 7, 3-bit; 0 to 15, 4-bit).
Tests are performed in the following order:
1. Limit Test 1
2. Limit Test 2
a. Lower Limit 2
b. Upper Limit 2
3. Limit Test x, where x = 3, 5-12 in ascending
numerical order.
a. Lower Limit x
b. Upper Limit x
The first failure in the test sequence determines the bit
pattern for the digital output port. Subsequent failures
in the test sequence will not change the defined digital
output pattern. NOTE that the output value can be
specified as a binary, octal, decimal, or hexadecimal
value.
Use the following table to determine the parameter
value for the desired decimal digital output pattern.
For non-decimal parameters, convert the decimal value
to its binary, octal, or hexadecimal equivalent.
Содержание GSM-20H10
Страница 107: ...GSM 20H10 User Manual 105 The operating flow of Grading mode is shown in the figure below ...
Страница 108: ...BASIC OPERATION 106 ...
Страница 110: ...BASIC OPERATION 108 ...
Страница 166: ...SYSTEM SETTINGS 164 ...
Страница 244: ...COMMAND SYNTAX 242 Example ROUTe TERMinals ...