COMMAND SYNTAX
224
Command :CALCulate3:FORMat <name>
Function
This Command is used to select the desired statistic on
readings stored in the buffer.
Readings stored in the buffer can be “raw” measured
readings, the results of the CALC1 calculation, or
CALC2 readings. The :TRACe:FEED command in
the :TRACe Subsystem is used to select the type of
readings to store.
<name> = MEAN Mean value of readings in
buffer
SDEViation Standard deviation of
readings in buffer
MAXimum Largest reading in buffer
MINimum Lowest reading in buffer
PKPK MAXimum - MINimum
Example
:CALCulate3:FORMat MAXimum
Command :CALCulate3:FORMat?
Function
Query programmed math format.
Example
:CALCulate3:FORMat?
Command :CALCulate3:DATA?
Function
This query Command is used to perform the selected
statistic operation and read the result(s). The result(s)
is always returned in ASCII format.
If the buffer is configured to store raw measured
readings (:TRACe:FEED SENSe1) and multiple
functions were measured, the selected statistic
operation will be performed on all the measured
readings.