APPENDIX
368
Programmed Source
Resolution
10pA, ±1.00000uA range
100pA, ±10.0000uA range
1nA, ±100.000uA range
10nA, ±1.00000mA range
100nA, ±10.00000mA range
1uA, ±100.000mA range
10uA, ±1.00000A range
Programmed Source
Accuracy
1
±(0.029%+300pA)), ±1.00000uA range
±(0.027%+700pA), ±10.0000uA range
±(0.025%+6nA), ±100.000uA range
±(0.027%+60nA), ±1.00000mA range
±(0.035%+600nA), ±10.00000mA range
±(0.055%+6uA), ±100.000mA range
±(0.22%+570uA), ±1.00000A range
Temperature
Coefficient (0°–18°C &
28°–50°C
±(0.1 × accuracy specification) / °C
Measurement
Resistance
Range
Resolution Test
current
Accuracy
<2.00000Ω ---
---
Source IACC+Meas.VACC
2.00000Ω
10uΩ
---
Source IACC+Meas.VACC
20.0000Ω
100uΩ
100mA ±(0.1%+0.003 Ω), NormaL
±(0.07%+0.001 Ω), Enhanced
200.000Ω
1mΩ
10mA
±(0.08%+0.03 Ω), NormaL
±(0.05%+0.01 Ω), Enhanced
2.00000kΩ 10mΩ
1mA
±(0.07%+0.3 Ω), NormaL
±(0.05%+0.1 Ω), Enhanced
20.0000kΩ 100mΩ
100uA
±(0.06%+3 Ω), NormaL
±(0.04%+1 Ω), Enhanced
200.000kΩ 1Ω
10uA
±(0.07%+30 Ω), NormaL
±(0.05%+10 Ω), Enhanced
2.00000MΩ 10Ω
5uA
±(0.11%+300 Ω), NormaL
±(0.05%+100 Ω), Enhanced
20.0000MΩ 100Ω
0.5uA
±(0.11%+1k Ω), NormaL
±(0.05%+500 Ω), Enhanced
200.000MΩ 1kΩ
100nA
±(0.66%+10k Ω), NormaL
±(0.35%+5k Ω), Enhanced
>200.000M
Ω
---
---
Source IACC+Meas.VACC
Temperature
Coefficient (0°–18°C &
28°–50°C)
±(0.15 × accuracy specification)/°C
Source I mode,Manual
OHMS
Total uncertainty = I source accuracy
+ V measure accuracy (4-wire remote
sense).
Source V mode, Manual
OHMS
Total uncertainty = V source accuracy
+ I measure accuracy (4-wire remote
sense).