GSM-20H10 User Manual
115
External I/O
Description The GSM can be connected to an external device
through the external DIGITAL I/O port. The
DIGITAL I/O port includes 4 output lines and 2 input
lines.
Pin
Definition
line5
:
+3.3V
line7
:
GND
line6,8
:
IDLE
line9: +5V output, used to drive external logic circuits.
The maximum output current of this port is 300mA.
line10: Start-of-test line, SOT is the input signal.
line11: Output enable, OE is the input signal. It is
used with an output enable circuit on a device or test
fixture.
line12-line15: 4 output signal ports of Digtal I/O ports
Connection
Diagram
The measurement method shown in the figure above
is used to send the pattern values of the limit test.
Load
connection
Each open-collector output can be set to high level
(+5V) or low level (0V), and can source 2mA current
or sink 500mA current.