The test passes when all readings measure less than the stated limits.
The actual location of the test probe may vary by system. Measurements should be recorded for each probe
under the set of test conditions.
Record all data and keep the record of the results with other hard copies of maintenance data.
Table 10-17 Typical Data Format for recording Probe (source) Leakage Current
Unit under Test:
Date of Test:
Test Conditions
Probe as measured in Saline Bath
System Power
Grounding/PE
4C-D
11L-D
RIC5-9-D
RM6C
off
closed
off
open
on
closed
on
open
Note
Values in italics font are given as examples only.
10.6.8.3 Isolated Probe (Sink) Leakage-Isolation Test
Warning
Do not use the probe if the insulating material has been punctured or otherwise compromised. Integrity of the
insulation material and patient safety can be verified by safety testing according to IEC60601-1.
Caution
Line voltage is applied to the ultrasound probe during this test. To avoid possible electric shock hazard, the
system being tested must not be touched by patients, users or anyone while the ISO TEST switch is
depressed.
Measurements should be recorded for full lead combination under each set of test conditions specified in:
•
Record all data and keep the record of the results with other hard copies of maintenance data.
1.
Connect the Safety analyzer to an AC wall outlet.
2.
Plug the equipment under test into the receptacle on the panel of the meter.
3.
Connect the ultrasound probe to be tested to the Voluson E-Series system.
4.
Immerse the saline probe in the saline solution.
5.
Immerse the Voluson E-Series probe´s face (imaging area of the probe) into the saline solution.
6.
Select the "Patient Lead Leakage" function on the meter.
7.
Test with closed ground with the Voluson E-Series system on and off.
Note
For more information, refer to the safety analyzer's user manual that will be used to perform the test.
Record all data and keep the record of the results with other hard copies of maintenance data.
Table 10-18 Typical Data Format for recording Isolated Probe (sink) Leakage
Unit under Test:
Date of Test:
Test Conditions
Probe as measured in Saline Bath
System Power
Grounding/PE
4C-D
11L-D
RIC5-9-D
RM6C
off
closed
on
closed
Note
Values in italics font are given as examples only.
Care and Maintenance
Voluson E-Series Service Manual
KTD106657 Revision 2
10-21
Содержание H48681XB
Страница 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 172: ...This page was intentionally left blank Functional Checks 4 46 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 226: ...This page was intentionally left blank Service Adjustments 6 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 288: ...This page was intentionally left blank Replacement Procedures 8 30 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 364: ...This page was intentionally left blank Care and Maintenance 10 24 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 365: ......
Страница 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...