4.3.9 Calculations
The Voluson E-Series system supports calculation packages and application-oriented patient Worksheets
(Reports) for following applications:
•
Abdomen Calculations
•
Small Parts Calculations
•
Obstetric Calculations
•
Cardiology Calculations
•
Urology Calculations
•
Vascular Calculations
•
Gynecology Calculations
•
Pediatric Calculations
•
Neurology Calculations
•
Orthopedics Calculations
Note
Confirm that the patient information is correct and the probe and application are selected properly.
Different menus are displayed depending on:
• the currently selected Application,
• the selected Display Mode,
• the selected “Study”,
• and the settings in the Measure Setup -
Measure & Calc
page.
Note
For further details refer to the Voluson E-Series Basic User Manual.
General remarks to perform Calculations:
•
By pressing the Calc key on the control console the Calculation function is switched on.
•
Positioning of measurement marks is done with the Trackball.
•
Entering and storage of measuring marks is done with Set (right or left trackball key).
•
To change measuring marks before completion press Change (upper trackball key).
•
Depending on the setting in the Measure Setup, also the Freeze key can be used for confirming the last measuring
mark of the currently performed measurement.
•
The status bar area shows the current function of the trackball.
•
To cancel the measurement of the currently selected item, touch
Cancel on the Touch Panel.
•
To delete the results of the last measured item, touch
Delete Last on the Touch Panel.
•
To delete all measurement results of the selected “Study” from the monitor as well as from the corresponding
Worksheet, touch the
Clear Study key on the Touch Panel.
•
All measurement results will be automatically included in the corresponding patient worksheet.
•
To erase measurement results from the screen, press the Clear key on the control console or press the Del key on
the keyboard.
•
To exit from Calculations touch the Exit key on the Touch Panel, press the Calc key or the Exit key on the control
console.
4.3.9.1 Worksheet (Report) Pages
Press the Report key on the control console to view the “application dependent” patient worksheet pages that contain the
results of calculations and measurements. Any stored patient worksheet can be edited, printed, transferred, saved in the
Archive or sent to DICOM server.
Functional Checks
Voluson E-Series Service Manual
KTD106657 Revision 2
4-25
Содержание H48681XB
Страница 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 172: ...This page was intentionally left blank Functional Checks 4 46 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 226: ...This page was intentionally left blank Service Adjustments 6 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 288: ...This page was intentionally left blank Replacement Procedures 8 30 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 364: ...This page was intentionally left blank Care and Maintenance 10 24 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 365: ......
Страница 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...