7.2 Request for Service (RFS)
Note
Service Connectivity has to be checked out once before you can request for service. i.e., Service Platform
has to be configured properly; see
Section 3.13.4 "InSite ExC Configuration" on page 3-77
There are 2 possibilities to contact GE:
•
by means of the GE “Remote Status Icon” that is displayed on the bottom of the screen
Move the cursor to the InSite ExC link (GE icon) at the right bottom of the display
screen and press the left trackball key (= left-click). The “Contact GE” form (see
•
via the System Setup "Service" page
1.
Press the Utilities key on the control console.
2.
In the “Utilities” menu touch the
System Setup
button to invoke the setup desktop on the screen.
3.
On the left side of the screen select
Administration
and then click the
Service
tab.
4.
Enter the password <SHE> and click the
Accept
button to display the Service Tools page (see
).
5.
Click the
RFS
button and fill out the displayed form. (Enter detailed problem description.)
Figure 7-3 Contact GE - Request for Service
Note
Connection must be checked out! (1)
6.
Click the
Send
button to send the problem description to GE Service/Application representatives.
Diagnostics/Troubleshooting
Voluson E-Series Service Manual
KTD106657 Revision 2
7-5
Содержание H48681XB
Страница 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 172: ...This page was intentionally left blank Functional Checks 4 46 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 226: ...This page was intentionally left blank Service Adjustments 6 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 288: ...This page was intentionally left blank Replacement Procedures 8 30 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 364: ...This page was intentionally left blank Care and Maintenance 10 24 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 365: ......
Страница 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...